FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING

MDR report key: 1952660 · Received January 3, 2011

Report

Report Number
2531779-2011-00021
Event Type
Malfunction
Date Received
January 3, 2011
Report Date
December 6, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED, THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE BOLUS BUTTON WAS DIFFICULT TO PUSH AND INTERMITTENTLY RESPONSIVE TO USER INPUT DURING TESTING, ALL KEYPAD BUTTONS WERE RESPONSIVE AND SPRUNG BACK WHEN PRESSED, AND IT WAS OBSERVED THAT THE BOLUS BUTTON KEY CONTACT ACTIVATOR WAS ROTATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE AUDIO BOLUS BUTTON IS INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. THE PTS CLAIMED THAT THE KEYPAD BUTTON STILL SPRING UP AND DOWN AND CLICK NORMALLY. THE PT DENIED TEARS OR PEELING OF THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1