FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1952640
·
Received January 3, 2011
Report
- Report Number
- 2531779-2011-00020
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Report Date
- December 6, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED, THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, ALL KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUT WHEN PRESSED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP, DOWN AND OK KEYPAD BUTTONS REPORTEDLY SLOW TO RESPOND WHEN PRESSED. THE REPORTER CLAIMED THAT THE KEYPAD BUTTONS SPRING BACK AS NORMAL AND THAT THE KEYPAD WAS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |