PENTAX
Report
- Report Number
- 9610877-2024-54488
- Event Type
- Malfunction
- Date Received
- June 13, 2024
- Date of Event
- June 5, 2024
- Report Date
- June 13, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- PMA / PMN Number
- K172156
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD SIGNAL WIRE RUPTURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD SIGNAL WIRE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LIGHT GUIDE CABLE COATING DAMAGE, THE LIGHT GUIDE CABLE FOR CONTROL BODY COATING DAMAGE, THE LIGHT GUIDE CABLE FOR PRONG COATING DAMAGE, THE SEGMENT BROKEN, THE INSERTION FLEXIBLE TUBE PERFORATED, THE REMOTE CONTROL BUTTONS PERFORATED, AND THE BENDING RUBBER STRETCHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 794274 | PENTAX | VIDEO NASO PHARYNGO LARYNGOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | VNL8-J10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |