AUTOSOFT 90
Report
- Report Number
- 3003442380-2024-07651
- Event Type
- Malfunction
- Date Received
- June 12, 2024
- Date of Event
- March 3, 2024
- Report Date
- February 10, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244014541
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1885107. DEVICE 1 OF 2.
SUPPLEMENTAL REPORT 01 - (B)(4). ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY THE COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6002813 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 1 AND WORK INSTRUCTION GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 1 FOR THE CODE "SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE" PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. THE FOLLOWING TEST WAS PERFORMED: VISUAL TEST ACCORDING TO WORK INSTRUCTION VERSION 1 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST 1 ACCORDING TO WORK INSTRUCTION VERSION 1 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. A BATCH RECORD REVIEW WAS CONDUCTED RESULTING IN THE FOLLOWING: THE LOT 6002813 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION VERSION 68 MANUFACTURED IN THE LINE 5, ON 15/AUG/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESS HAD BEEN FULFILLED AND MET THE REQUIREMENTS. A QUERY WAS RUN IN DATABASE AGAINST MALFUNCTION CODE IDD-PMC05.47 AND LOT 6002813 AND OTHER ZERO COMPLAINT HAS BEEN REGISTERED IN DATABASE SINCE THE LAST 12 MONTHS. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: HARM NO REPORTABLE, NO DEFECT ON TESTS, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, OTHER ZERO COMPLAINTS RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVELLIENCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE MARKET QUALITY REVIEW (MQR) PROCEDURE.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. BETWEEN (B)(6) 2024, IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA KINKED EVENTS. THE EVENTS OCCURRED WITHIN 3 HOURS OF INSERTION. THE INSERTION SITE WAS AT THE ABDOMEN. THE BLOOD GLUCOSE LEVEL WAS 332 MG/DL AT THE TIME OF EVENT. THE PATIENT REPLACED THE INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1184167 | AUTOSOFT 90 | INSET II 2-PACK 60/6 GREY TCAP | FPA | UNOMEDICAL A/S | 1000281 | 6002813 | 05705244014541 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female |