AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-07634
- Event Type
- Malfunction
- Date Received
- June 12, 2024
- Date of Event
- March 1, 2024
- Report Date
- June 12, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4). DEVICE 2 OF 2.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON 01-MAR-2024, IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA BENT EVENTS. THE EVENTS OCCURRED WITHIN 3 HOURS OF INSERTION. THE INSERTION SITE WAS AT THE ABDOMEN. THE BLOOD GLUCOSE LEVEL WAS 160 - 320 MG/DL AT THE TIME OF EVENT. THE PATIENT REPLACED THE INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 937731 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001680 | UNKNOWN | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR | Male |