CONTAK RENEWAL
Report
- Report Number
- 2124215-2010-23350
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- December 6, 2010
- Report Date
- December 16, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ANALYSIS IS CURRENTLY ON-GOING. UPON COMPLETION OF ANALYSIS THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED AND REPLACED FOR NORMAL BATTERY DEPLETION. THERE WERE NO FIELD ALLEGATIONS REPORTED AGAINST THIS PRODUCT. THIS REPORT WAS CREATED DUE TO LABORATORY FINDINGS AS THIS DEVICE DID NOT PASSED LONGEVITY CALCULATIONS. SHORTENED REPLACEMENT WINDOW (4/05/2007).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H179 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | 0154| H179| E110| MISMATCH| 1388T| 1861 |