FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1952520 · Received January 7, 2011

Report

Report Number
2124215-2010-23332
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
December 6, 2010
Report Date
December 10, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DETAILED ANALYSIS THIS EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY VOLTAGE WAS 2.32 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO ONE COMPROMISED LOW-VOLTAGE CAPACITOR, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECIEVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED WITH NO ALLEGATIONS. INTIAL DEVICE ANALYSIS SHOWED THAT THE DEVICE FAILED LONGEVITY. ADDITIONAL TESTING WILL BE CONDUCTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1