FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1952506 · Received January 7, 2011

Report

Report Number
2124215-2010-23455
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
December 6, 2010
Report Date
December 9, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1620-09 TO Z-1621-09
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS DEVICE IS PART OF THE SHORTENED REPLACEMENT WINDOW ADVISORY, ORIGINALLY COMMUNICATED (B)(4), 2007.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS REPORT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. INITIAL LABORATORY ANALYSIS FOUND THE DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H175

Patients

Seq Age Sex Outcome Treatment
1 81 YR H175| N118| 0158| 4470| 4518| 4525