VITALITY 2
Report
- Report Number
- 2124215-2010-23305
- Event Type
- Injury
- Date Received
- January 7, 2011
- Date of Event
- December 3, 2010
- Report Date
- January 3, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-1620-09 TO Z-1621-09
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF 2.45 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION.
THE ICD WAS SUCCESSFULLY REPLACED AND WILL BE RETURNED FOR LABORATORY ANALYSIS. NO ADDITIONAL INFORMATION IS AVAILABLE. ONCE THE DEVICE HAS BEEN RETURNED AND ANALYSIS COMPLETED, THIS REPORT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AFTER EXPERIENCING PREMATURE BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
THE ICD WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |