FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 1952431
·
Received January 3, 2011
Report
- Report Number
- 1824206-2011-00012
- Event Type
- Malfunction
- Date Received
- January 3, 2011
- Date of Event
- December 13, 2010
- Report Date
- December 13, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND FAULTY BRAKE CASTERS AND REPLACED ALL BRAKE CASTERS AND BOTH HEX RODS TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
TECHNICIAN ALLEGED THAT ALL BRAKE CASTERS SWIVEL WHEN IN THE BRAKE POSITION. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM INC. | 8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |