FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 19524087 · Received June 12, 2024

Report

Report Number
2024168-2024-07019
Event Type
Injury
Date Received
June 12, 2024
Date of Event
May 22, 2024
Report Date
October 2, 2024
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648344039
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTIONS: D1 ¿ BRAND NAME: UPDATED D2A ¿ COMMON DEVICE NAME: UPDATED D3 ¿ NAME, ADDRESS 1, CITY, POSTAL CODE: UPDATED

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. D4- THE UDI IS NOT KNOWN AS THE PART AND LOT NUMBER WERE NOT PROVIDED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE REPORTED INFORMATION AND RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. CORRECTIONS: D4 - LOT #: UPDATED FROM UNKNOWN TO 4021441 D4 - CATALOG #: UPDATED FROM UNK PROSTYLE TO 12773-02 D4 - PRIMARY UDI NUMBER: UPDATED FROM UNKNOWN TO (B)(6). D9 - DEVICE AVAILABLE FOR EVALUATION: UPDATED FROM "YES" TO "NO".

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VENOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING A PROSTYLE DEVICE AFTER AN ELECTROPHYSIOLOGY (EP) INTERVENTIONAL PROCEDURE USING A SMALL BORE SHEATH. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] WAS NOTICED. ANOTHER PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
161530 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-02 4021441 08717648344039

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention