FDA Adverse Event Injury Summary report: N

EASYTRAK 3

MDR report key: 1952290 · Received January 7, 2011

Report

Report Number
2124215-2010-23750
Event Type
Injury
Date Received
January 7, 2011
Date of Event
December 2, 2010
Report Date
January 5, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THIS PRODUCT HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

THIS PRODUCT IS CURRENTLY UNDERGOING DETAILED ANALYSIS. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

THE LEAD WAS LATER RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS FOUND TO BE DISLODGED. A REVISION PROCEDURE WAS PERFORMED, DURING WHICH THIS LEAD WAS EXPLANTED TO IMPLANT A NEW LEAD WITHOUT BLOOD IN THE LUMEN. TO DATE, THERE HAVE BEEN NO REPORTED ADVERSE PATIENT EFFECTS RELATED TO THIS CLINICAL OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 3 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4548

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention 4136| 4543| 4137| 4592| 4548| H120