FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 1952271 · Received January 7, 2011

Report

Report Number
2124215-2010-24115
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
November 29, 2010
Report Date
January 21, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. ANALYSIS OF THE ASSOCIATED LEADS CONCLUDED THE EVENT OCCURRED AS A RESULT OF IMPROPER INSERTION OF THE LEAD INTO THE DEVICE HEADER.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT, DURING AN IMPLANT PROCEDURE, NOISE WAS NOTED ON THE ELECTROGRAM OF THIS DEVICE AFTER LEAD CONNECTIONS WERE COMPLETED. SENSING MEASUREMENTS COULD NOT BE COMPLETED DUE TO THE NOISE. AFTER SEVERAL ATTEMPTS TO RECONNECT THE LEADS, NOISE WAS STILL SENSED. ALSO, THE DEVICE MEASURED PACING IMPEDANCES GREATER THAN 2,000 OHMS ON BOTH THE LEFT VENTRICULAR (LV) AND RIGHT VENTRICULAR (RV) CHANNELS. THE LEAD CONNECTIONS WERE CHECKED AGAIN SEVERAL TIMES. THE RV AND LV LEAD IMPEDANCES WERE MEASURED WITH A PACING SYSTEM ANALYZER, AND WERE FOUND TO BE IN THE NORMAL RANGE. TROUBLESHOOTING STEPS WERE TAKEN IN AN ATTEMPT TO REDUCE EXTERNAL NOISE SOURCES, BUT THE PROBLEM COULD NOT BE RESOLVED. THE DEVICE WAS REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND P108

Patients

Seq Age Sex Outcome Treatment
1