FDA Adverse Event Malfunction Summary report: N

TELIGEN

MDR report key: 1952267 · Received January 7, 2011

Report

Report Number
2124215-2010-23957
Event Type
Malfunction
Date Received
January 7, 2011
Date of Event
October 28, 2010
Report Date
August 13, 2018
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS ICD REMAINS IMPLANTED, SO THEREFORE WILL NOT BE RETURNED TO BOSTON SCIENTIFIC FOR LABORATORY ANALYSIS. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED LOW PACING IMPEDANCE MEASUREMENTS. THIS PATIENT WILL BE FOLLOWED-UP IN THE NEAR FUTURE TO EVALUATE RV PACING LEAD. SUBSEQUENTLY, ADDITIONAL INFORMATION WAS RECEIVED THAT A LOW SHOCK IMPEDANCE WAS ALSO OBSERVED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Description of Event or Problem · 1

NEW INFORMATION RECEIVED INDICATES THAT THIS DEVICE WAS EXPLANTED ALMOST 7 YEARS LATER FOR AN UNKNOWN REASON AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND F110

Patients

Seq Age Sex Outcome Treatment
1 73 YR