IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM
Report
- Report Number
- 0002023141-2024-01946
- Event Type
- Malfunction
- Date Received
- June 12, 2024
- Date of Event
- May 8, 2024
- Report Date
- October 22, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024020054
- PMA / PMN Number
- K061410
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4109, 4111 AND 3331. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 213. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 67. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED IMPLANT FOR EVALUATION. VISUAL EVALUATION / FUNCTIONAL TEST WAS PERFORMED AND COVER SCREW DETACHED FROM IMPLANT AS INTENDED. DAMAGED DUE TO THE REMOVAL PROCESS. NO MALFUNCTION IDENTIFIED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1231838. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1231838 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL: FUNCTIONAL: DOES NOT DISENGAGE/RELEASE A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED SINCE DEVICE MALFUNCTION DID NOT OCCUR, AND THE REPORTED EVENT HAS BEEN UNCONFIRMED FOLLOWING FUNCTIONAL TESTING AND PHYSICAL EVALUATION. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID NOT OCCUR. NO MALFUNCTION DISCOVERED. IMPLANT DETACHED FROM MOUNT AS INTENDED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A4: WEIGHT UNKNOWN / NOT PROVIDED G4: ADDITIONAL PMA/510(K) NUMBER ¿ K013227 PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
IT WAS REPORTED THAT UPON THE TORQUE TEST APPOINTMENT FOR THE IMPLANT AT TOOTH SITE #20, THE IMPLANT PASSED TEST BUT THE COVER SCREW WAS COLD WELDED. THE IMPLANT STARTED TO TURN WHEN THEY TRIED TO REMOVE THE SCREW. THE CLINICIAN HAD TO EXTRACT THE IMPLANT AND REPLACE IT WITH A NEW ONE. SYMPTOMS AS A RESULT OF THE EVENT: PAIN
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182522 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1231838 | 00889024020054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Male |