FDA Adverse Event Malfunction Summary report: N

SYRINGE 20ML LL S/C 48

MDR report key: 19522047 · Received June 12, 2024

Report

Report Number
1911916-2024-00419
Event Type
Malfunction
Date Received
June 12, 2024
Date of Event
June 3, 2024
Report Date
June 21, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903028307
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP IT WAS REPORTED THE PLUNGER BROKE WHEN ASPIRATING AIR. AS A SAMPLE WAS NOT RETURNED, A THOROUGH SAMPLE EVALUATION COULD NOT BE PERFORMED. TO AID IN THE INVESTIGATION, ONE PHOTO AND TWO VIDEOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE MEDIA SHOWS A NEEDLE ASSEMBLED TO A SYRINGE THAT IS DRAWING UP A DRUG. WHEN PULLING THE PLUNGER ROD UP A DAMAGED PLUNGER ROD RIB IS OBSERVED. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THERE WAS A JAM DURING THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302830, LOT 2242123. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. VERIFICATION OF THE ASSEMBLY PROCESS WAS PERFORMED. THE ALIGNMENT OF THE RAILS AND CONVEYORS WAS CORRECT. THE FLOW OF PRODUCT WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Description of Event or Problem · 0

NEW 20ML SYRINGE PLUNGER BROKE WHEN THE ASPIRATING AIR IN THE PHARMACY.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED IMPACTED LOT NO. 2242123, EXP: 31-08-2027 NEW 20ML SYRINGE PLUNGER BROKE WHEN THE ASPIRATING AIR IN THE PHARMACY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188878 SYRINGE 20ML LL S/C 48 SYRINGE, PISTON FMF BECTON DICKINSON 2242123 30382903028307

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown