FDA Adverse Event Injury Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1952109 · Received December 23, 2010

Report

Report Number
2531779-2010-02910
Event Type
Injury
Date Received
December 23, 2010
Date of Event
November 5, 2010
Report Date
November 17, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT BEING RETURNED TO ANIMAS FOR EVAL. NO MECHANICAL ISSUES WERE OBSERVED THROUGH TROUBLESHOOTING.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: REVIEW OF THE BLACK BOX AND DOWNLOAD HISTORY DID NOT CONTAIN DATA FROM THE DATE OF THE ALLEGED INCIDENT, (B)(4) 2012, DUE TO CONTINUED PATIENT USE. DATA WAS AVAILABLE FOR (B)(4) 2012 THRU (B)(4) 2012. A REVIEW OF THE TOTAL DAILY DOSE HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THE PUMP WAS EXERCISED FOR 29 HOURS WITH NO DELIVERY ISSUES. THE PUMP WAS EVALUATED AND FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING INSULIN ACCURATELY.

Description of Event or Problem · 1

THE RPTR CONTACTED ANIMAS ON (B)(6) 2010, ALLEGING HIGH BLOOD GLUCOSE LEVELS. THE RPTR CONTACTED ANIMAS ON BEHALF OF HER SON (THE PT). SINCE (B)(6) 2010, THE RPTR CLAIMED THAT THE PT'S BLOOD GLUCOSE LEVELS HAD BEEN AROUND 200-600 MG/DL. THE RPTR CLAIMED THAT A DOCTOR HAD THE PT COME OFF OF THE PUMP AND GO TO INJECTIONS. THE RPTR ALSO CLAIMED THAT THE PT'S BLOOD GLUCOSE LEVEL HAD COME DOWN TO THE 180 MG/DL RANGE. THE PT WAS REPORTEDLY CHANGING SITES EVERY 2-3 DAY. NO BENDING OR KINKING OF THE CANNULAS WERE OBSERVED. NO SIGNS OF INSULIN LEAKING WERE ALSO OBSERVED. THE RPTR DID NOT HAVE CARTRIDGE WITH THE PUMP FOR TROUBLESHOOTING. THE RPTR DENIED THAT THE PT HAD ANY RECENT ILLNESSES OR NEW MEDICATIONS. THE RPTR WAS ADVISED TO TRY USING AN INFUSION SET FROM A NEW BOX.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 11 YR Life Threatening