FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1952083 · Received January 4, 2011

Report

Report Number
1720753-2011-00027
Event Type
Malfunction
Date Received
January 4, 2011
Date of Event
December 15, 2010
Report Date
January 4, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND RESEATED CONNECTORS ON THE BACK PLANE. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THE SYSTEM DISPLAYED A COMMUNICATION ERROR AND WOULD NOT FLUORO. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1