ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-23216
- Event Type
- Injury
- Date Received
- January 7, 2011
- Date of Event
- November 29, 2010
- Report Date
- December 14, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. SET SCREW MARKS WERE NOTED ON THE IS-1 TERMINAL PIN, AND THE DISTAL AND PROXIMAL HIGH VOLTAGE TERMINAL PINS. NO DISCERNABLE SET SCREW MARKS WERE NOTED ON THE IS-1 RING. MULTIPLE CUTS WERE NOTED IN THE LEAD INSULATION ALONG WITH BLOOD/BODY FLUID NOTED IN THE LUMENS DUE TO THE CUTS. THERE WAS NO SUTURE SLEEVE RETURNED ON THE LEAD. ELECTRICAL RESISTANCE TESTING OF THE CONDUCTOR PATHS SHOWED THE LEAD TO BE IN SPECIFICATION. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING DEFIBRILLATION THRESHOLD (DFT) TESTING WITH THIS IMPLANTABLE RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD A FAULT MESSAGE WAS RECEIVED INDICATING AN OPEN CIRCUIT DETECTED. THE TERMINAL PIN WAS VERIFIED TO BE IN THE PROPER PORT OF THE DEVICE HEADER AND THE SHOCKING IMPEDANCE MEASUREMENTS WERE IN RANGE AT 56 OHMS. THE POCKET WAS REOPENED AND THE LEAD WAS EXPLANTED. THERE HAD ALSO BEEN REPORTS OF HIGH DFTS EVEN AFTER LEAD REPOSITIONING. A NEW LEAD WAS IMPLANTED AND THE ISSUE RESOLVED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Required Intervention | 0181| E110| 4135| 0185 |