FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 1952077 · Received January 7, 2011

Report

Report Number
2124215-2010-23216
Event Type
Injury
Date Received
January 7, 2011
Date of Event
November 29, 2010
Report Date
December 14, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. SET SCREW MARKS WERE NOTED ON THE IS-1 TERMINAL PIN, AND THE DISTAL AND PROXIMAL HIGH VOLTAGE TERMINAL PINS. NO DISCERNABLE SET SCREW MARKS WERE NOTED ON THE IS-1 RING. MULTIPLE CUTS WERE NOTED IN THE LEAD INSULATION ALONG WITH BLOOD/BODY FLUID NOTED IN THE LUMENS DUE TO THE CUTS. THERE WAS NO SUTURE SLEEVE RETURNED ON THE LEAD. ELECTRICAL RESISTANCE TESTING OF THE CONDUCTOR PATHS SHOWED THE LEAD TO BE IN SPECIFICATION. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING DEFIBRILLATION THRESHOLD (DFT) TESTING WITH THIS IMPLANTABLE RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD A FAULT MESSAGE WAS RECEIVED INDICATING AN OPEN CIRCUIT DETECTED. THE TERMINAL PIN WAS VERIFIED TO BE IN THE PROPER PORT OF THE DEVICE HEADER AND THE SHOCKING IMPEDANCE MEASUREMENTS WERE IN RANGE AT 56 OHMS. THE POCKET WAS REOPENED AND THE LEAD WAS EXPLANTED. THERE HAD ALSO BEEN REPORTS OF HIGH DFTS EVEN AFTER LEAD REPOSITIONING. A NEW LEAD WAS IMPLANTED AND THE ISSUE RESOLVED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0181

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention 0181| E110| 4135| 0185