FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1952021 · Received January 7, 2011

Report

Report Number
2124215-2010-23244
Event Type
Injury
Date Received
January 7, 2011
Date of Event
December 1, 2010
Report Date
January 7, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-0047-08 TO Z-0053-08
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, TECHNICIANS UTILIZED AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATE THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED TO VERIFY THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING EXCEEDED THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE THAT THIS DEVICE IS FAULTY, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, OR IF THE DEVICE IS RETURNED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO EXTENDED CHARGE TIMES EXCEEDING THE EXTENDED CHARGE TIME LIMIT. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention 1821| T165| MISMATCH| 0125