FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1952017 · Received January 7, 2011

Report

Report Number
2124215-2010-23236
Event Type
Injury
Date Received
January 7, 2011
Date of Event
December 2, 2010
Report Date
January 3, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1620-09 TO Z-1621-09
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF 2.22 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Additional Manufacturer Narrative · 1

THE ICD WAS SUCCESSFULLY REPLACED AND WILL BE RETURNED FOR LABORATORY ANALYSIS. NO ADDITIONAL INFORMATION IS AVAILABLE. ONCE THE DEVICE HAS BEEN RETURNED AND ANALYSIS COMPLETED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED THE ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER BEING IMPLANTED FOR 54 MONTHS. THERE WAS A CONCERN THAT THE BATTERY HAD DEPLETED EARLIER THAN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention T165