FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1951898 · Received December 29, 2010

Report

Report Number
3003464075-2010-00149
Event Type
Injury
Date Received
December 29, 2010
Date of Event
December 1, 2010
Report Date
December 1, 2010
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK AFTER DIALYSATE WAS DEPLETED DURING TREATMENT. FACILITY STAFF HAS BEEN NOTIFIED. THERE HAVE BEEN NO SIMILAR PROBLEMS REPORTED SUBSEQUENT TO THIS EVENT. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Description of Event or Problem · 1

DIALYSATE DEPLETED PRIOR TO COMPLETION OF A ROUTINE HEMODIALYSIS TREATMENT. THE OPERATOR DISCONTINUED TREATMENT WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. THE PT'S STANDARD EPOGEN DOSE WAS INCREASED. NO OTHER MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170-B 0097705

Patients

Seq Age Sex Outcome Treatment
1 47 YR Other