UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-00085
- Event Type
- Malfunction
- Date Received
- January 7, 2011
- Date of Event
- December 8, 2010
- Report Date
- December 8, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER HAD LOADED FRESH ISE REAGENTS, CALIBRATED AND QC WAS WITHIN LAB-ESTABLISHED RANGES. WHEN THE CUSTOMER NOTICED HIGH NA RESULTS, QC WAS RERUN AND FAILED LAB-ESTABLISHED RANGES. SERVICE WAS ALREADY ON-SITE FOR A DIFFERENT ISSUE, AND FOUND THAT THE ISE REAGENT TUBING WAS PINCHED BY THE REAGENT DOOR, WHICH WAS RESTRICTING THE REAGENT FLOW. THE ISSUE HAD OCCURRED WHEN THE CUSTOMER LOADED THE REAGENT AND CLOSED THE REAGENT DOOR. THE FIELD SERVICE ENGINEER (FSE) RESOLVED THE TUBING ROUTING ISSUE, WHICH WAS THE ROOT CAUSE OF THIS ISSUE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS OBTAINING FALSE HIGH SODIUM (NA), CHLORIDE (CL), CARBON DIOXIDE (CO2) RESULTS, AND FALSE LOW CALCIUM (CALC) RESULTS THAT WERE GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. ONE POTASSIUM (K) SAMPLE EXCEEDED ASSAY PRECISION BY 1.0 MMOL/L. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. AFTER HIGH NA RESULTS WERE NOTICED, AND TROUBLESHOOTING WAS PERFORMED, SAMPLES WERE RERUN AND REPORTS WERE AMENDED. A TOTAL OF 31 REPORTS WERE CORRECTED AND 28 REPORTS WERE CORRECTED BEFORE THE FALSE RESULTS REACHED THE PHYSICIAN. TWO EXAMPLES OF THE ERRONEOUS RESULTS AND THE RERUN RESULTS, PROVIDED BY THE CUSTOMER. TREATMENT WAS NOT INITIATED OR WITHHELD BASED UPON THE FALSE RESULTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JGS | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |