FDA Adverse Event
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 1951760
·
Received January 6, 2011
Report
- Report Number
- 2050012-2011-00067
- Date Received
- January 6, 2011
- Date of Event
- December 10, 2010
- Report Date
- December 10, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) FOUND THE CAP PIERCER BLADE BROKEN AND SHAVED PIECES OFF THE BLADE WICK AND CLOGGED THE BLADE WASH DRAIN. THE CUSTOMER TECHNICAL SUPPORT (CTS) BLADE COUNT WAS OVER 87000. THE CUSTOMER ORDERED A REPLACEMENT BLADE. FSE INSTRUCTED THE CUSTOMER ON HOW TO CHANGE THE BLADE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT FLUID LEAK FROM THE CAP PIERCER BLADE WASH STATION. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |