FDA Adverse Event Summary report: N

UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1951760 · Received January 6, 2011

Report

Report Number
2050012-2011-00067
Date Received
January 6, 2011
Date of Event
December 10, 2010
Report Date
December 10, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) FOUND THE CAP PIERCER BLADE BROKEN AND SHAVED PIECES OFF THE BLADE WICK AND CLOGGED THE BLADE WASH DRAIN. THE CUSTOMER TECHNICAL SUPPORT (CTS) BLADE COUNT WAS OVER 87000. THE CUSTOMER ORDERED A REPLACEMENT BLADE. FSE INSTRUCTED THE CUSTOMER ON HOW TO CHANGE THE BLADE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT FLUID LEAK FROM THE CAP PIERCER BLADE WASH STATION. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1