FDA Adverse Event Malfunction Summary report: N

INTEGRATED APD SET W/CASSETTE3-PRONG

MDR report key: 1951716 · Received January 6, 2011

Report

Report Number
1423500-2011-00210
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 10, 2010
Report Date
December 10, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE ROOT CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE SUPPLY BAG FELL AND DISCONNECTED. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) UNIT. THE HOME PATIENT (HP) WAS STILL CONNECTED. ONE OF THE SUPPLY BAGS FELL AND BECAME DISCONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HP AND ASSISTED TO CYCLE POWER TO THE HC MACHINE. THE HP WOULD FINISH WITH MANUALS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.

Description of Event or Problem · 1

DURING A FOLLOW UP WITH THE HOME PATIENT (HP), THE HP STATED THAT SHE WAS AT A HOTEL AND SHE MOVED THE LAMP AND THINGS ON THE NIGHTSTAND SO SHE COULD SET UP FOR THERAPY. THE HP STATED THAT THE NIGHT STAND WAS KIND OF TALL SO SHE COULD NOT PUT THE BAG ON THE FLOOR SO SHE PUT THE CYCLER ON THE NIGHTSTAND AND PUT THE SOLUTION BAG BEHIND IT. THE HP STATED THAT THE ONLY THING SHE COULD THINK OF IS THAT WHEN THE CYCLER WAS PULLING SOLUTION FROM THE SUPPLY BAG, IT PULLED ENOUGH FOR THE BAG TO FALL AND THEN THE BAG BECAME DISCONNECTED. THE HP STATED THAT SHE STARTED OVER WITH NEW SUPPLIES TO FINISH THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTEGRATED APD SET W/CASSETTE3-PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HC CYCLER