INTEGRATED APD SET W/CASSETTE3-PRONG
Report
- Report Number
- 1423500-2011-00210
- Event Type
- Malfunction
- Date Received
- January 6, 2011
- Date of Event
- December 10, 2010
- Report Date
- December 10, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE ROOT CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE SUPPLY BAG FELL AND DISCONNECTED. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) UNIT. THE HOME PATIENT (HP) WAS STILL CONNECTED. ONE OF THE SUPPLY BAGS FELL AND BECAME DISCONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HP AND ASSISTED TO CYCLE POWER TO THE HC MACHINE. THE HP WOULD FINISH WITH MANUALS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.
DURING A FOLLOW UP WITH THE HOME PATIENT (HP), THE HP STATED THAT SHE WAS AT A HOTEL AND SHE MOVED THE LAMP AND THINGS ON THE NIGHTSTAND SO SHE COULD SET UP FOR THERAPY. THE HP STATED THAT THE NIGHT STAND WAS KIND OF TALL SO SHE COULD NOT PUT THE BAG ON THE FLOOR SO SHE PUT THE CYCLER ON THE NIGHTSTAND AND PUT THE SOLUTION BAG BEHIND IT. THE HP STATED THAT THE ONLY THING SHE COULD THINK OF IS THAT WHEN THE CYCLER WAS PULLING SOLUTION FROM THE SUPPLY BAG, IT PULLED ENOUGH FOR THE BAG TO FALL AND THEN THE BAG BECAME DISCONNECTED. THE HP STATED THAT SHE STARTED OVER WITH NEW SUPPLIES TO FINISH THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | HC CYCLER |