FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 1951676 · Received January 6, 2011

Report

Report Number
2024168-2011-00079
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 13, 2010
Report Date
December 14, 2010
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
P810046
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURES INCLUDE, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, WEAK MATERIALS, EXCESSIVELY APPLIED PRESSURE, OR INTERACTIONS WITH ACCESSORY DEVICES, PATIENT ANATOMY, AND/OR LESION CALCIFICATION OR TORTUOSITY. RETURN OF THE VOYAGER NC CATHETER USED IN THE PROCEDURE MAY HAVE AIDED IN THE INVESTIGATION AND DETERMINATION OF CAUSE. THERE WAS NO REPORT OF ANY LEAK IN THE CATHETER NOTED DURING PREPARATION FOR USE, WHICH WOULD SUGGEST THAT THE BALLOON WAS NOT DAMAGED PRIOR TO USE. IT IS POSSIBLE THAT THE BALLOON MATERIAL WAS DAMAGED (SCRATCHED) DURING INTERACTIONS WITH ACCESSORY DEVICES OR THE MODERATELY CALCIFIED PATIENT ANATOMY, SUCH THAT THE BALLOON RUPTURED UPON THE SECOND INFLATION AT 14 ATM WHICH IS BELOW THE RATED BURST PRESSURE (RBP) OF 18 ATM. A REVIEW OF THE PRODUCT MANUFACTURING RECORDS DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. THERE IS NO INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE CANNOT BE DETERMINED. TO ENSURE THIS IS NOT A RESULT OF MANUFACTURING, ALL BALLOON CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED PRIOR TO PACKAGING. ADDITIONALLY, A SAMPLING OF UNITS IS DESTRUCTIVELY TESTED TO VERIFY RBP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN INTERVENTIONAL PROCEDURE OF A CONCENTRIC, MODERATELY TORTUOUS AND CALCIFIED, DE NOVO, DISTAL CIRCUMFLEX LESION, THE 3.0 X 18 MM VOYAGER NC WAS USED FOR PRE-DILATATION DURING A FIRST INFLATION AT 10 ATMOSPHERES (ATM); HOWEVER, THE BALLOON RUPTURED DURING THE SECOND INFLATION AT 14 ATM. THE PROCEDURE WAS COMPLETED USING A 3.0 X 11 MM NON-ABBOTT BALLOON. THERE WAS NO REPORTED PATIENT EFFECT. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER NC CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 0033161

Patients

Seq Age Sex Outcome Treatment
1 DILATATION CATHETER: 3.0 X 11 MM MERCURY