ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-23607
- Event Type
- Injury
- Date Received
- January 6, 2011
- Date of Event
- November 27, 2010
- Report Date
- November 27, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THE EVENT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE, RIGHT VENTRICULAR (RV) LEAD AND LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH PACE IMPEDANCE MEASUREMENTS GREATER THAN 2,000 OHMS. THE PATIENT WAS BROUGHT INTO THE OFFICE FOR TROUBLESHOOTING AND IMPEDANCE WAS WITHIN RANGE, WITH AND WITHOUT ISOMETRICS. DATA IN LATITUDE REVEALED THAT BOTH RV AND LV IMPEDANCE HAVE BEEN GREATER THAN 2,000 OHMS ON EVERY DAILY MEASUREMENT FOR APPROXIMATELY THE PAST 2 WEEKS. THERE WERE EPISODES DISPLAYING NOISE ON THE RV CHANNEL. THE POCKET WAS OPENED FOR FURTHER EVALUATION. THE PHYSICIAN TUGGED ON THE RV LEAD AND THE CONNECTION WAS SECURE, HOWEVER, IT APPEARED THAT THE LEAD MAY NOT HAVE BEEN ADVANCED FULLY INTO THE HEADER. THE PHYSICIAN DISCONNECTED THE RV LEAD FROM THE DEVICE AND TESTED THE LEAD WITH THE PACING SYSTEM ANALYZER. THE LEAD WAS RECONNECTED TO THE DEVICE AND THE PHYSICIAN CONFIRMED THAT A SECURE CONNECTION HAD BEEN ESTABLISHED. ALL DIAGNOSTICS WERE WITHIN RANGE FOLLOWING RECONNECTION. MULTIPLE IMPEDANCE TESTS WERE PERFORMED WITH POCKET MANIPULATION AND ALL REVEALED NORMAL IMPEDANCE MEASUREMENTS. NO FURTHER IMPEDANCE ISSUES HAVE BEEN REPORTED. ALL AVAILABLE INFORMATION INDICATES THAT THE DEVICE AND LEADS REMAIN IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0158 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention | N119| 1488T| H170| 0158| 4542 |