FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1951476 · Received January 6, 2011

Report

Report Number
2124215-2010-23096
Event Type
Injury
Date Received
January 6, 2011
Date of Event
November 25, 2010
Report Date
December 16, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE ICD WAS ANALYZED. A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

THE ICD WAS SUCCESSFULLY REPLACED AND WILL BE RETURNED FOR LABORATORY ANALYSIS. NO ADDITIONAL INFORMATION IS AVAILABLE. ONCE THE DEVICE HAS BEEN RETURNED AND ANALYSIS COMPLETED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), IMPLANTED FOR 41 MONTHS, DISPLAYED THE ELECTIVE REPLACEMENT INDICATOR (ERI). THE CHARGE TIME WAS 18.6 SECONDS WITH A MONITORING VOLTAGE OF 2.64 VOLTS. THERE WAS A CONCERN THAT THE BATTERY HAD DEPLETED EARLIER THAN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1