LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3002158293-2010-01306
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- November 19, 2010
- Report Date
- December 6, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE BATTERY CHARGER WAS NOT FUNCTIONAL AND WOULD NOT CHARGE THE PT'S BATTERY PACKS. UPON EVALUATION, THE POWER BRICK WAS FOUND TO BE DEFECTIVE. THE ROOT CAUSE OF THE DEFECTIVE POWER BRICK CANNOT BE POSITIVELY DETERMINED. ONCE THE POWER BRICK WAS REPLACED, THE CHARGER WAS FULLY FUNCTIONAL. NO ADVERSE EVENT RESULTED FROM THE INTERMITTENT BATTERY CHARGER. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER.
A PT SERVICE REPRESENTATIVE ASSISTING A (B)(6) FEMALE PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT SERVICE TO REPORT AN ISSUE WITH A CHARGER. SHE REPORTS THE CHARGER WOULD NOT POWER ON, DESPITE PLUGGING IT INTO MULTIPLE OUTLETS. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY CHARGER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |