FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 1951214 · Received January 6, 2011

Report

Report Number
2954323-2011-00079
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
November 18, 2010
Report Date
January 6, 2011
Product Code
NBW
Removal / Correction Number
ADC FA1197-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30 DEGREES CELSIUS DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D, OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. CUSTOMERS WILL BE NOTIFIED THROUGH A REMEDIAL ACTION ADC FA1197-2010 LETTER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THE TEST DID NOT START WHEN THE TEST STRIP WAS INSERTED INTO THEIR ADC BLOOD GLUCOSE METER AFTER BLOOD WAS APPLIED. CUSTOMER REPORTED USING AN AFFECTED TEST STRIP LOT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.