FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 1951185 · Received December 14, 2010

Report

Report Number
1824206-2010-11572
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
November 18, 2010
Report Date
November 18, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE HEAD HI/LOW CYLINDER WAS WORN. HE REPLACED THE HEAD HI/LOW CYLINDER TO REPAIR THE STRETCHER.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD HI/LOW FUNCTION IS DRIFTING DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1