FDA Adverse Event Malfunction Summary report: N

(NONE FOUND)

MDR report key: 1951106 · Received December 29, 2010

Report

Report Number
MW5018841
Event Type
Malfunction
Date Received
December 29, 2010
Report Date
December 28, 2010
Manufacturer
CLINTON INDUSTRIES
Product Code
KMN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THIS DEVICE IS A SAFETY BLOOD DRAW CHAIR, IN THAT AN ARM IS PROVIDED TO KEEP THE PT FROM FALLING FORWARD, IF PT PASSES OUT DURING BLOOD DRAW PROCESS. THE PROBLEM IS THE NARROW BASE OF THE CHAIR; IF ONE FALLS FORWARD, THE ENTIRE CHAIR GOES WITH THEM, ENDING UP ON TOP OF THEM AS THE FALL TO THE FLOOR. WE HAD TO ATTACH THE CHAIR TO THE WALL TO MAKE IT SAFE. THE MANUFACTURER AND SELLER BOTH IGNORED US WHEN WE REPORTED THE PROBLEM TO THEM SOME TIME AGO. THE SAME ITEM STILL APPEARS IN THEIR NEW CATALOG. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 (NONE FOUND) EXTRA-TALL UPHOLSTERED BLOOD DRAW CHAIR KMN CLINTON INDUSTRIES NONE GIVEN NONE GIVEN

Patients

Seq Age Sex Outcome Treatment
1