FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 1951052 · Received January 6, 2011

Report

Report Number
2124215-2010-23408
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
October 26, 2010
Report Date
November 19, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

TH DEVICE WAS SUCCESSFULLY INTERROGATED BY BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY. THE PACING, SENSING, AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION FROM AN IMPLANT FORM THAT THIS DEVICE, WHICH WAS IMPLANTED FOR 58.5 MONTHS, WAS ELECTIVELY EXPLANTED AND REPLACED DUE TO NORMAL BATTERY DEPLETION. END-OF-LIFE (EOL) WAS TRIGGERED IN (B)(6) 2010 WITH A MONITORING VOLTAGE OF 2.68 VOLTS ASSOCIATED WITH A CHARGE TIME OF 33.5 SECONDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T180

Patients

Seq Age Sex Outcome Treatment
1 59 YR (B)(4)| (B)(4)| (B)(4)