FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 1951004 · Received January 6, 2011

Report

Report Number
2124215-2010-23258
Event Type
Injury
Date Received
January 6, 2011
Date of Event
November 22, 2010
Report Date
March 11, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THE LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, IT WAS NOTED THAT THE COMPLETE LEAD WAS RETURNED WITH A SET SCREW MARK ON THE TERMINAL PIN AND RING. VISUAL INSPECTION FOUND TISSUE IN THE HELIX MECHANISM. FURTHER DETAILED ANALYSIS REVEALED THAT THE OUTER INSULATION HAD RUPTURED THROUGH ON BOTH SIDES AT ABOUT 26.8 CM TO 28.2 CM FROM THE TERMINAL PIN. THE DAMAGED INSULATION HAS A ZIPPER-LIKE APPEARANCE AND THE OUTER COILS UNDERNEATH WERE SLIGHTLY FLATTENED. THE FIELD ALLEGATION OF DISLODGEMENT WAS NOT CONFIRMED. HOWEVER, ANALYSIS DID FIND INSULATION DAMAGE WITH THE COILS EXPOSED; THIS TYPE OF DAMAGE AND LOCATION (26.8 CM TO 28.2 CM FROM PIN) IS CONSISTENT WITH CLAVICLE FIRST/RIB ENTRAPMENT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE EXPLANT PROCEDURE FOR THE NON-BOSTON SCIENTIFIC RIGHT ATRIAL LEAD DISLODGEMENT, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE, A DISLODGEMENT OR EXIT BLOCK ARE SUSPECTED. A TEMPORARY PACER WAS PLACED AND BOTH LEADS WERE EXPLANTED. NEW RA AND RV LEADS WERE SUCCESSFULLY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4088

Patients

Seq Age Sex Outcome Treatment
1 44 YR Hospitalization| L| R S601| 5180| (B)(4)| MISMATCH| 4088| 4471