FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1950992 · Received January 3, 2011

Report

Report Number
3004464228-2010-01472
Event Type
Malfunction
Date Received
January 3, 2011
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE RETURNED POD FOUND EVIDENCE OF AN INTERNAL FLUID LEAK. DISCOLORATION WAS SEEN INSIDE THE DEVICE AND RESIDUAL FLUID AND CORROSION WAS FOUND ON THE SURFACES OF INTERNAL ASSEMBLIES. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS.

Description of Event or Problem · 1

THE REPORT INDICATED THAT THE CUSTOMER EXPERIENCED SMALL KETONES AND CONSISTENTLY HIGH BG LEVELS (345-486MG/DL) THAT WOULD NOT LOWER DESPITE MULTIPLE CORRECTION BOLUS ATTEMPTS. THE CUSTOMER'S MOTHER REPORTED THAT THE POD HAD INITIATED AN ALARM (THE SPECIFIC ALARM TYPE WAS NOT PROVIDED) AND STATED THAT "THE CANNULA LOOKED BENT" WHEN THE POD WAS REMOVED. THE POD WILL BE RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30393

Patients

Seq Age Sex Outcome Treatment
1 13 YR Other