FDA Adverse Event Malfunction Summary report: N

DUCKBILL

MDR report key: 1950972 · Received January 6, 2011

Report

Report Number
1219602-2011-00001
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 1, 2010
Report Date
December 6, 2010
Manufacturer
SMITH & NEPHEW ENDOSCOPY
Product Code
NBH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION OF THE DEVICE SHOWS THE INNER TIP IS MISSING AND WAS NOT RETURNED FOR EVALUATION. THE CUTTING EDGE IS DENTED AND DULL. A PORTION OF THE ACTUATOR LUG IS MISSING. THE CONDITION OF THE CUTTING EDGE WOULD REQUIRE EXCESSIVE FORCE BE APPLIED TO CUT. AS REPORTED DEVICE WAS PURCHASED IN 2002 AND REPAIRED THE LAST TIME IN 2006. DEVICE IS IN NEED OR REFURBISHMENT/MAINTENANCE AND MOST LIKELY DID AT THE TIME OF THIS EVENT.(B)(4)

Description of Event or Problem · 1

THE DUCKBILL BROKE DURING THE SURGERY AND THE SURGEON HAD TO OPEN THE KNEE TO REMOVE THE BROKEN PIECE (OPEN ON 4 CM); BUT THE INCIDENT OCCURRED AT THE END OF THE SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUCKBILL PCH,BSKT,DCKBL,UPBTR,1.5MM NBH SMITH & NEPHEW ENDOSCOPY 012014 L602

Patients

Seq Age Sex Outcome Treatment
1