FDA Adverse Event
Malfunction
Summary report: N
DEXTRUS 4136
MDR report key: 1950914
·
Received November 4, 2010
Report
- Report Number
- 1028232-2010-02292
- Event Type
- Malfunction
- Date Received
- November 4, 2010
- Date of Event
- August 23, 2010
- Report Date
- October 5, 2010
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
Description of Event or Problem · 1
BOSTON SCIENTIFIC CRM RECEIVED INFO THAT THIS RV LEAD WAS SURGICALLY ABANDONED DUE TO A LOW IMPEDANCE MEASUREMENT OF 180 OHMS AND INSULATION DAMAGE. A NEW RV LEAD WAS SUCCESSFULLY IMPLANTED. NO ADVERSE PT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4136 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |