SYSMEX AUTOMATED COAGULATION ANALYZER CA-1500
Report
- Report Number
- 1422681-2010-00008
- Event Type
- Malfunction
- Date Received
- August 9, 2010
- Date of Event
- April 6, 2010
- Report Date
- June 16, 2010
- Manufacturer
- SYSMEX CORP
- Product Code
- JPA
- PMA / PMN Number
- K011235
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE MFR SYSMEX CORP (B)(4) PERFORMED INVESTIGATION AND CORRECTIVE ACTION. ROOT CAUSE INVESTIGATION: OBSERVATION OF DEFECTIVE PARTS REVEALED THAT A CRACK WAS GENERATED AT THE DRIVING PART OF THE SAMPLE TUBE. THE INVESTIGATION CONFIRMED THAT THERE IS NOT ANY PROBLEM IN QUALITY OF THE TUBING MATERIAL AND IN THE MFG PROCESS. HOWEVER, THE DRIVING PART OF THE SAMPLE TUBE MAY CONTACT A BUR AND/ OR ROUGH EDGE GENERATED DURING SPOT WELDING ON THE FRAME. CONSIDERING ITS MOVEMENT, THE CONTACT IS RARE. IN CONCLUSION, THE LIKELY CAUSE FOR THE MICRO-HOLE (CRACK) IN THE SAMPLE ARM TUBING WAS CAUSED BY THE MOVEMENT OF ITS DRIVING PART IN LONG TERM USE, PROBABLY BY CONTACT WITH SOME BURR AND ROUGH EDGE ON THE FRAME. FCA TO BE TAKEN IN THE NEAR FUTURE: THE SAMPLE ARM TUBING WILL BE COVERED BY A "SLEEVE" TO PROTECT IT FROM MECHANICAL VULNERABILITY. AUGUST 2010 THROUGH OCTOBER 2010; ADDITIONAL INFO-PREVIOUS COUNTERMEASURES PERFORMED: RELEASED THE GUIDELINES "PROCEDURE HOW TO EXCHANGE TUBES ON CA-1500" JUNE 2008; EXECUTED DESIGN CHANGE REGARDING THE ROUTING OF THE TUBING IN OUR MFG PROCESS. OCTOBER 2008; SHIPPED 1,000 CA-1500S FOR OVERSEAS AND 70 FOR (B)(4) MARKET AFTER DESIGN CHANGE. NO COMPLAINTS HAVE BEEN RECEIVED REGARDING A MICRO HOLE IN THE CA-1500S AFTER THE DESIGN CHANGE. METHOD: THE INVESTIGATION INTO ROOT CAUSE WAS PERFORMED BY (B)(4). CONCLUSION: FIELD CORRECTION BY (B)(4).
RANDOMLY INCORRECT ELEVATED PT AND APTT RESULTS WERE REPORTED ON (B)(6) 2010, CAUSED BY LEAKAGE FROM THE CA-1500 SAMPLE ARM TUBING IN THE ANALYZER. ONE PT (HOSPITAL IDENTIFIER: PT 1-FEMALE) HAD A POSSIBLE CHANGE TO THEIR DOSAGE OF WARFARIN BASED ON THE ERRONEOUS RESULT. TWO PTS (HOSPITAL IDENTIFIERS: PT 2-MALE AND PT 6-FEMALE) RECEIVED INCORRECT MEDICAL TREATMENT (TRANSFUSION OF BLOOD PRODUCTS) AS A RESULT OF THE INCORRECT REPORTS. NO ADDITIONAL PT IMPACT WAS REPORTED. PT 6-FEMALE: WAS TESTED AT 715 PM ON (B)(6) 2010. THIS PT WAS GIVEN FOUR UNITS OF FRESH FROZEN PLASMA (FFP) AND 4 UNITS OF CRYOPRECIPITATE (CRYO). REPEAT TESTING OF THE SAME SAMPLE GAVE NORMAL RESULTS; THEREFORE THE 8 UNITS OF BLOOD PRODUCT ADMINISTERED WAS UNNECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYSMEX AUTOMATED COAGULATION ANALYZER CA-1500 | CA-1500 | JPA | SYSMEX CORP | CA-1500 W/CP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |