FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 1950771 · Received January 6, 2011

Report

Report Number
2124215-2010-22617
Event Type
Injury
Date Received
January 6, 2011
Date of Event
November 18, 2010
Report Date
December 6, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION RECEIVED INDICATED THE PATIENT WAS BROUGHT IN AND THE NOISE WAS UNABLE TO REPRODUCED WITH POCKET MANIPULATIONS. A REVISION PROCEDURE TOOK PLACE AND THE CONNECTIONS WERE VERIFIED. THE PACE/SENSE PORTION OF THE RV LEAD WAS SURGICALLY ABANDONED AND A NEW PACE/SENSE LEAD WAS IMPLANTED. THE CHRONIC RV LEAD REMAINS IN SERVICE FOR DEFIBRILLATION. A HIGH VOLTAGE SHOCK WAS DELIVERED AND ALL MEASUREMENTS WERE WITHIN NORMAL RANGE. ALL AVAILABLE INFORMATION INDICATES THAT THE DEVICE REMAINS IN SERVICE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION HAVE BEEN MADE. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD WERE EXHIBITING NOISE ON THE RATE/SENSE AND SHOCK CHANNEL. THIS PATIENT IS PACEMAKER DEPENDANT AND PACING WAS INHIBITED FOR APPROXIMATELY 5 SECONDS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE PATIENT PLANNED TO BE SENT TO THEIR DEVICE FOLLOWING PHYSICIAN FOR A PROBABLE LEAD REPLACEMENT. THERE WERE ALSO SOME DIVERTED EPISODES DUE TO THE INAPPROPRIATE SENSING, HOWEVER NO SHOCKS HAD BEEN DELIVERED. THE NOISE WAS ABLE TO BE REPRODUCED WITH ARM MOVEMENTS. THERE HAD BEEN NO CHANGE IN IMPEDANCES, THRESHOLDS, OR SENSING MEASUREMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0185

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention 0185| MISMATCH| 4469| (B)(4)| 4542| H210| 1290