FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1950755 · Received January 6, 2011

Report

Report Number
2124215-2010-22421
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
November 18, 2010
Report Date
November 23, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE, WHICH WAS INCLUDED IN THE SUBPECTORAL IMPLANT (B)(6) 2008 POPULATION PRODUCT ADVISORY WAS INITIALLY COMMUNICATED ON (B)(6) 2006, WAS EXPLANTED FOR AN UNSPECIFIED REASON. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED WITH THIS CLINICAL OBSERVATION. THE DEVICE WAS RETURNED FOR RELIABILITY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 70 YR T177| 0185