FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 ARTICULATING

MDR report key: 1950694 · Received January 6, 2011

Report

Report Number
1527736-2011-00009
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 3, 2010
Report Date
December 13, 2010
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCOMPLETE/INTERRUPTED FIRING CYCLE. THE EC60A DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH A CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED. A PARTIAL FIRE CAN OCCUR IF THE FIRING SEQUENCE IS INTERRUPTED, THE DEVICE IS OPENED, THEN CLOSED AND FIRING IS RESUMED, CAUSING THE INSTRUMENT TO LOCKOUT. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION USING A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A LOCKED DEVICE A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. ONCE THE DEVICE IS OPENED THE DEVICE CAN BE DE-ARTICULATED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 ARTICULATING STAPLE, IMPLANTABLE GDW ETHICON ENDO SURGERY, INC (CINCINNATI) UNK G4U12H

Patients

Seq Age Sex Outcome Treatment
1