FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 1950521 · Received January 6, 2011

Report

Report Number
2124215-2010-22775
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
November 17, 2010
Report Date
November 17, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT REMAINS IN SERVICE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS TRANSVENOUS DEFIBRILLATION LEAD EXHIBITED EPISODES OF NOISE ON THE RIGHT VENTRICULAR (RV) CHANNEL, WHICH WAS OVERSENSED AND RESULTED IN 3 TO 5 SECOND PAUSES IN PACING. THERE WAS A SMALL AMOUNT OF NOISE ON THE SHOCK CHANNEL. NORMAL PACING VALUES WERE OTHERWISE NOTED. DEVICE MEMORY WAS SAVED TO A DISK AND RETURNED. ENGINEERING REVIEW OF THE DEVICE MEMORY NOTED OVERSENSING OF EXTERNAL NOISE AND THE PRESENCE OF A MAGNET OVER THE DEVICE AT THE TIME OF THE EPISODES. THEY ADVISED TO INQUIRE IF THE PATIENT HAD A PROCEDURE THAT DAY, AS THE NOISE MIGHT BE DUE TO ELECTROCAUTERY. IT WAS ALSO NOTED THAT THE SHOCK CHANNEL RECORDED WHAT APPEARED TO BE PATIENT VENTRICULAR FIBRILLATION (VF) OR PAROXYSMAL VENTRICULAR TACHYCARDIA (PVT). IT WAS THEN REPORTED THAT THERE HAD BEEN EPISODES OF NOISE ON THE RV CHANNEL WITH ISOMETRICS PRIOR TO THIS; HOWEVER, THERE WERE NO EPISODES IN THE ARRHYTHMIA LOGBOOK. IT WAS NOTED THAT THE PATIENT WOULD BE BROUGHT BACK IN TO THE CLINIC FOR FURTHER EVALUATION. TO DATE, THERE HAVE BEEN NO REPORTED ADVERSE PATIENT EFFECTS RELATED TO THIS CLINICAL OBSERVATION. ADDITIONAL INFORMATION WAS PROVIDED THAT THE PATIENT WAS HAVING A SKIN CANCER REMOVAL WHEN THE PAUSES OCCURRED. THUS, EXTERNAL NOISE WAS THE CAUSE OF THE INHIBITION OF PACING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0184

Patients

Seq Age Sex Outcome Treatment
1 84 YR MISMATCH| N119| 0184| 1488TC| 4555