FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1950464 · Received January 6, 2011

Report

Report Number
2124215-2010-22420
Event Type
Injury
Date Received
January 6, 2011
Date of Event
November 15, 2010
Report Date
November 29, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-1513-08 TO Z-1514-08
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND DETAILED ANALYSIS IS PENDING.

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

A RETURN REQUEST WAS MADE FOR THIS PRODUCT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD DECLARED THE ELECTIVE REPLACEMENT INDICATOR (ERI) WITH A CHARGE TIME OF 19.1 SECONDS. ON THE DAY OF EXPLANT, THE DEVICE THEN MEASURED 2.67 VOLTS WITH CHARGE TIMES OF 16.6 SECONDS. TECHNICAL SERVICES DISCUSSED DEVICE BEHAVIOR AND RECOMMENDED REPLACEMENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS ADVISORY.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention 0158| T177