FDA Adverse Event Malfunction Summary report: N

COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L1-TE4BE1AB

MDR report key: 1950314 · Received December 14, 2010

Report

Report Number
2030404-2010-00263
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
November 16, 2010
Report Date
November 16, 2010
Manufacturer
ST. JUDE MEDICAL
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE RETURNED CATHETER REVEALED NO ANOMALIES. FUNCTIONAL TESTING REVEALED THE CATHETER PASSED CONTINUITY, RESISTANCE AND EKG TESTING. STEERING FORCE TO CREATE A CURVE WAS WITHIN SPEC AT 9.5 LBS. THE CATHETER ALSO SUCCESSFULLY PASSED ABLATION SIMULATION TESTING, PRESSURE DROP TESTING AND THE RATE TEST. THE TIP OF THE CATHETER WAS VIEWED UNDER MICROSCOPE AND NO ANOMALIES WERE NOTED. TO SIMULATE THE COMPLAINT ISSUE THE ABLATION SIMULATION TEST WAS PERFORMED TWICE. DURING THE FIRST ABLATION SIMULATION TEST THE PUMP FLOW RATE WAS SET AT 17 ML/MIN AND THE GENERATOR WAS SET TO THE 50W PER LAB STANDARD. THE 2ND TEST WAS PERFORMED PER CUSTOMER SETTING OF 30 ML/MIN FLOW RATE ON THE PUMP AND GENERATOR WAS SET TO 40W. EACH TIME THE TEST WAS PERFORMED WITHOUT ANY ISSUE. THE CAUSE FOR THE REPORTED "POP" COULD NOT BE DETERMINED. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THIS DEVICE MET MFG REQUIREMENTS PRIOR TO SHIPMENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHILE USING THE CATHETER FOR RIGHT ATRIAL FLUTTER ISTHMUS ABLATION PROCEDURE, "POPPING" WAS NOTED DURING ABLATION AND ISTHMUS BLOCK COULD NOT BE ACHIEVED. THE IRRIGATION PUMP WAS SET AT 30 ML/MIN AND THE GENERATOR WAS SET AT 40 WATTS. THERE WERE NO CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L1-TE4BE1AB NONE OAD ST. JUDE MEDICAL 83565 K12594

Patients

Seq Age Sex Outcome Treatment
1 UNK