STRATA 2, ADJUSTABLE VALVE, REGULAR
Report
- Report Number
- 2021898-2010-00344
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 9, 2010
- Report Date
- November 10, 2010
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- PMA / PMN Number
- K042465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE VALVE WAS PATENT, BUT DID NOT MET THE REQUIREMENTS FOR LEAK TESTING DUE TO LARGE TEARS IN THE TOP THE DELTA CHAMBER. THIS PRECLUDED SIPHON AND REFLUX TESTING AS WELL AS MEASUREMENT OF PRESSURE FLOW CHARACTERISTIC AND PREIMPLANTATION TESTING. IT IS UNK HOW OR WHEN THESE DAMAGES OCCURRED. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT CARE SHOULD BE TAKEN IN HANDLING THE VALVES AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE. IT IS ALSO SUGGESTED THAT THE VALVE BE PLACED IN A SURGICALLY CREATED LOOSE SUBGALEAL POCKET. THIS ALLOWS GENTLE PLACEMENT OF THE VALVE AND MINIMIZES HANDLING WITH SURGICAL TOOLS. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. NO PT IMPACT WAS REPORTED. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MANUFACTURE.
IT WAS REPORTED THAT THE DELTA CHAMBER OF THE VALVE WAS BROKEN DURING IMPLANTATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRATA 2, ADJUSTABLE VALVE, REGULAR | JXG | MEDTRONIC NEUROSURGERY | NA | C66638 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |