FDA Adverse Event Malfunction Summary report: N

STRATA 2, ADJUSTABLE VALVE, REGULAR

MDR report key: 1950227 · Received December 9, 2010

Report

Report Number
2021898-2010-00344
Event Type
Malfunction
Date Received
December 9, 2010
Date of Event
November 9, 2010
Report Date
November 10, 2010
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K042465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE VALVE WAS PATENT, BUT DID NOT MET THE REQUIREMENTS FOR LEAK TESTING DUE TO LARGE TEARS IN THE TOP THE DELTA CHAMBER. THIS PRECLUDED SIPHON AND REFLUX TESTING AS WELL AS MEASUREMENT OF PRESSURE FLOW CHARACTERISTIC AND PREIMPLANTATION TESTING. IT IS UNK HOW OR WHEN THESE DAMAGES OCCURRED. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT CARE SHOULD BE TAKEN IN HANDLING THE VALVES AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE. IT IS ALSO SUGGESTED THAT THE VALVE BE PLACED IN A SURGICALLY CREATED LOOSE SUBGALEAL POCKET. THIS ALLOWS GENTLE PLACEMENT OF THE VALVE AND MINIMIZES HANDLING WITH SURGICAL TOOLS. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. NO PT IMPACT WAS REPORTED. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MANUFACTURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DELTA CHAMBER OF THE VALVE WAS BROKEN DURING IMPLANTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRATA 2, ADJUSTABLE VALVE, REGULAR JXG MEDTRONIC NEUROSURGERY NA C66638

Patients

Seq Age Sex Outcome Treatment
1 UNK