FDA Adverse Event Malfunction Summary report: N

PLMA DVC V11.51 1 N

MDR report key: 1950184 · Received December 9, 2010

Report

Report Number
2921482-2010-00969
Event Type
Malfunction
Date Received
December 9, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING A THE USER FACILITY, THE DEVICE WAS ABLE TO BE PROGRAMMED WHILE THE LOCKOUT SWITCH WAS IN THE LOCKED POSITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLMA DVC V11.51 1 N 80FRN FRN HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA (B)(4)| PLUM A+ SOFTWARE MODULE, LIST # 12097