FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE)

MDR report key: 1950095 · Received December 7, 2010

Report

Report Number
2024601-2010-00986
Event Type
Malfunction
Date Received
December 7, 2010
Date of Event
September 20, 2010
Report Date
November 9, 2010
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. MEDWATCH SENT TO FDA ON: (B)(4) 2010. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ALLERGAN HAS RECEIVED THE PRODUCT; HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING." "CAUTION: FAILURE TO CREATE A STABLE, SMOOTH PATH FOR THE ACCESS PORT TUBING, WITHOUT SHARP TURNS OR BENDS, CAN RESULT IN TUBING BREAKS AND LEAKAGE."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED A "PORT EXPLANT DUE TO LEAK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE) LTI ALLERGAN NA 1716402

Patients

Seq Age Sex Outcome Treatment
1 33 YR