FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1950036 · Received January 6, 2011

Report

Report Number
2250051-2011-00006
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 20, 2010
Report Date
January 6, 2011
Manufacturer
HAMILTON BONADUZ AG
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE FOUND THAT THE PLUNGER CLAMP AT POSITION #6 HAD A STUCK EJECTION PIN. THE FSE REPLACED THE PLUNGER CLAMP AND LLD SPRING IN POSITION 6. LLD ERRORS CONTINUED WHEN PROCESSING SPLLD CHECK IN DIAGNOSTIC SOFTWARE. FSE INSPECTED THE 2-POLE CABLE AT POSITION #6 AND FOUND A BREAK IN CABLE. THE FSE REPLACED THE 2-POLE CABLE. THE FSE RE-PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).