ORTHO SUMMIT SAMPLE HANDLING SYSTEM
Report
- Report Number
- 2250051-2011-00006
- Event Type
- Malfunction
- Date Received
- January 6, 2011
- Date of Event
- December 20, 2010
- Report Date
- January 6, 2011
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- JTC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE FOUND THAT THE PLUNGER CLAMP AT POSITION #6 HAD A STUCK EJECTION PIN. THE FSE REPLACED THE PLUNGER CLAMP AND LLD SPRING IN POSITION 6. LLD ERRORS CONTINUED WHEN PROCESSING SPLLD CHECK IN DIAGNOSTIC SOFTWARE. FSE INSPECTED THE 2-POLE CABLE AT POSITION #6 AND FOUND A BREAK IN CABLE. THE FSE REPLACED THE 2-POLE CABLE. THE FSE RE-PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE PIPETTER | JTC | HAMILTON BONADUZ AG | 936400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |