CLEARCUT KNIVES
Report
- Report Number
- 2523835-2024-01337
- Event Type
- Malfunction
- Date Received
- June 10, 2024
- Date of Event
- May 8, 2024
- Report Date
- September 19, 2024
- Manufacturer
- ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
- Product Code
- HNN
- UDI-DI
- 00000000000000
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CORRECTED INFORMATION IS PROVIDED IN D.4 AND ADDITIONAL INFORMATION IS PROVIDED IN SECTIONS D.9, H.3, H.6 AND H.11. ONE OPENED KNIFE, IN A BLISTER WAS RECEIVED FOR THE REPORT OF BLUNT KNIFE. SAMPLE WAS VISUALLY INSPECTED AND FOUND TO BE NONCONFORMING, BLADE WAS OBSERVED TO HAVE A BENT TIP AND A DAMAGED CUTTING EDGE. PENETRATION TESTING COULD NOT BE PERFORMED DUE TO THE DAMAGE OF THE SAMPLE. A REVIEW OF THE DEVICE HISTORY RECORD TRACEABLE TO THE REPORTED LOT NUMBERS, INDICATES THAT THE REPORTED PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE REPORTED PRODUCT¿S ACCEPTANCE CRITERIA THE EXACT ROOT CAUSE COULD NOT BE DETERMINED FROM THE INVESTIGATION PERFORMED. THE DAMAGE TO THE RETURNED SAMPLES IS CONSISTENT WITH DAMAGE THAT CAN OCCUR WHEN THE BLADE CONTACTS A HARD SURFACE SUCH AS THE PROTECTIVE BLADE TRAY WHEN PRODUCT IS IMPROPERLY REMOVED OR INSERTED AFTER USE, OR IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT DURING SURGERY OR SET-UP. THE DAMAGE SEEN ON THE RETURNED COMPLAINT SAMPLE COULD HAVE CONTRIBUTED TO THE CUSTOMERS REPORTED ISSUE OF BLUNT KNIFE. THE EXACT ROOT CAUSE FOR THE DAMAGED KNIFE SAMPLE IS UNKNOWN, THEREFORE SPECIFIC ACTION WITH REGARDS TO THIS COMPLAINT CANNOT BE TAKEN. A NONCONFORMANCE INVESTIGATION WAS COMPLETED TO INVESTIGATE THE TREND IDENTIFIED FOR DULL CUTTING INSTRUMENTS. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY NONCONFORMANCE, SUCH AS THE DAMAGED TIP AND DAMAGED CUTTING EDGE EXHIBITED ON THE RETURNED OPENED SAMPLE, IS REMOVED FROM THE LOT AND SCRAPPED. FUNCTIONAL PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. COMPLAINTS ARE REVIEWED AND MONITORED AT REGULAR INTERVALS FOR ANY SIGNIFICANT ADVERSE TRENDS. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). H.10 REFLECTS ALL RELATED REPORT NUMBERS ASSOCIATED WITH THIS PRODUCT EVENT THAT HAVE BEEN SUBMITTED AT THIS TIME.
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4). H.10 REFLECTS ALL RELATED REPORT NUMBERS ASSOCIATED WITH THIS PRODUCT EVENT THAT HAVE BEEN SUBMITTED AT THIS TIME.
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING CATARACT SURGERY IT WAS FOUND THAT THE KNIFE WAS DULL. SURGERY WAS COMPLETED WITH AN ALTERNATE BLADE. PATIENT IMPACT WERE NOT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 609403 | CLEARCUT KNIVES | KNIFE, OPHTHALMIC | HNN | ALCON RESEARCH, LLC - ALCON PRECISION DEVICE | NA | 167PW9 | 00000000000000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Male | CUSTOM-PAK SURGICAL PROCEDURE PACK |