PUMP MMT-1884L MM780G BLE 3.0 MG/ML
Report
- Report Number
- 2032227-2024-184309
- Event Type
- Malfunction
- Date Received
- June 10, 2024
- Date of Event
- May 20, 2024
- Report Date
- July 17, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- 003
Narratives
UNIT RECEIVED WITH CRITICAL PUMP ERROR (OPEN BOOK IMAGE). P-CAP LOCKED IN PLACE PROPERLY DURING TESTING. UNIT WAS SUCCESSFULLY DOWNLOADED USING THUMP SOFTWARE. UNABLE TO PERFORM THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, AND DISPLACEMENT TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK). THE ADAPT TOOL WAS UTILIZED TO SEARCH FOR ALARMS THAT MAY HAVE OCCURRED IN THE PAST OR DURING COMPLAIN CALL THAT MIGHT OF TRIGGER THE REASON COMPLAIN. DURING DOWNLOAD REVIEW USING ADAPT TOOL IT RECORDED PUMP ERROR 63 (MAIN ERROR LOG) FILE NUMBER = 2006, 49 LINE NUMBER = 704, 19825. PER SOFTWARE ENGINEER LOG IT WAS DETERMINED THE PUMP ERROR 63 WAS DUE TO HARDWARE ISSUE WITH LCD. UNIT WAS CUT OPEN AND PERFORM A VISUAL INSPECTION OF CONNECTORS AND ELECTRONIC ASSEMBLIES. PER VISUAL INSPECTION, IT WAS DETERMINED THAT THE INGRESS OF WATER CORRODING ELECTRONIC ASSEMBLIES AND FORCE SENSOR CAUSING ELECTRICAL SHORT. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CRACKED LCD WINDOW (MINOR), CRACKED KEYPAD OVERLAY AND SCRATCHED CASE. IN CONCLUSION, CUSTOMER CONCERN FOR CRITICAL PUMP ERROR (OPEN BOOK IMAGE) WAS TRIGGERED BY MOISTURE DAMAGE CONFIRMED AT THE ELECTRONIC ASSEMBLIES. PUMP ERROR 63 WAS CONFIRMED DUE TO POSSIBLE HARDWARE ISSUE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED CRITICAL PUMP ERROR (OPEN BOOK IMAGE). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1884L. TROUBLESHOOTING WAS PERFORMED AND CONFIRMED THE CUSTOMER RECEIVED CRITICAL PUMP ERROR/OPEN BOOK IMAGE ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1884L WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED. CUSTOMER WILL DISCONTINUE USING THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148513 | PUMP MMT-1884L MM780G BLE 3.0 MG/ML | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1884L | NG3537243H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Male |